FDA Recall Terminated

ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142

Recall: Z-0043-2008 · Initiated September 20, 2007

Recall

Recall Number
Z-0043-2008
Event Number
44909
Firm
Esa Biosciences, Inc.
FEI Number
1218996
Product Code
DOF
Status
Terminated
Root Cause
Process design
Initiated
September 20, 2007
Posted
October 18, 2007
Terminated
March 25, 2008
Address
22 Alpha Rd, Chelmsford, MA, 01824-4123

Description

ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142

Reason

Level 1 and Level 2 Control values were incorrectly assigned.

Action

ESA notifed Distributors and End Users by telephone and letter on 9/20/07. Distributors and End Users will be informed of the Level 1 and 2 Control value assignment change. End Users will be sent a new Level 1 and 2 Control Vial label for each vial they originally received. Distributors will be asked to return all test kits in their inventory to ESA Biosciences.

Distribution

Nationwide and United Kingdom.

Quantity

2 kits