FDA Recall Terminated

Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***.

Recall: Z-0043-04 · Initiated September 23, 2003

Recall

Recall Number
Z-0043-04
Event Number
27301
Firm
Microaire Surgical Instruments
FEI Number
2020601
Status
Terminated
Root Cause
Other
Initiated
September 23, 2003
Posted
October 15, 2003
Terminated
January 21, 2004
Address
1641 Edlich Dr, Charlottesville, VA, 22911

Description

Wire Driver Coupler, Model 7100-045, a component of MicroAire Drill/Reamer Model 7500. Driver couplers are reuseable medical devices, non-sterile, packed one per box and labeled in part ***MicroAire Surgical Instruments, 1641 Edlich Drive, Charlottesville, VA 22911***.

Reason

Wiring pin locking mechanism defective, potential to come apart in surgical site.

Action

The firm notified consignees by certified letter 9/23/03. The recall notifications were serial number specific and advised consignees of the potential defect with the wirning pin/locking mechanism. Consignees were instructed to cease use and distribution and obtain a return authorization for repair. Distributors were requested to extend the recall and recover and return devices for repair.

Distribution

Product was distributed to direct accounts around the United States and to distributors around the world.

Quantity

135