NxTAG Respiratory Pathogen Panel, REF: I051C0447
Recall
- Recall Number
- Z-0042-2025
- Event Number
- 95377
- Firm
- Luminex Molecular Diagnostics Inc
- FEI Number
- 3002777243
- Product Code
- OCC
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- September 5, 2024
- Posted
- October 9, 2024
- Address
- 439 University Ave, Toronto Canada Suite 900
Description
NxTAG Respiratory Pathogen Panel, REF: I051C0447
Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.
On 9/5/24, recall notices were sent to customers and distributors were were asked to do the following: 1. Two options are available to overcome the sporadic potential risk of the hMPV false positive result: A. Continue to use the product as per its instruction for use: in case of a positive result for hMPV, consider performing confirmatory testing with another method before reporting the hMPV result, if medically indicated. B. Exclude the hMPV from the reporting by removing this target from the assay analysis, as explained in the package insert under the chapter "Define Test Panels in SYNCT Software". See Document No. and Revision/Page reference: MLD-051-KPI-001 Rev H/Page 17, MLD-056-KPI-001 Rev. E/Page 18. 2. Notify others within your network that may have remaining inventory of those kits. 3. If you are a distributor, provide a copy of this letter to customers or organizations who received impacted lots. 4. Complete and return the response form via email to [email protected] 5. For complaints, questions, or concerns, please contact Technical Support using the contact information: 1-877-785-2323 (U.S. and Canada). +1-512-381-4397 (Outside U.S. and Canada). 1-512-219-5114 (Fax). [email protected] www.diasorin.com
U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.