Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
Recall
- Recall Number
- Z-0042-2020
- Event Number
- 83467
- Firm
- Roche Diagnostics Operations, Inc.
- FEI Number
- 1823260
- Product Code
- LEH
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- July 23, 2019
- Terminated
- June 8, 2020
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
Inaccurately Low Sample Results
1. Consignees will be provided instruction to download the updated e-barcode either from the cobas link or from the cobas e-library when available. 2. If analyzers cobas link is not connected to the Roche network, a future version of the cobas e-library CD will contain the revised e-barcodes. Download the e-barcode when available. 3. If analyzers cobas link is not connected to the Roche network and they require the updated VANC3 application e-barcode before the cobas e-library CD is available, they will be instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336 to request a Field Application Specialist to provide the VANC3 updated application. 4. Calibrate and run quality controls (QC) before running samples. 5.Complete all sections of the enclosed fax form. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
AK AL AR AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY
27676 units