FDA Recall Terminated

Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Recall: Z-0041-2020 · Initiated October 5, 2018

Recall

Recall Number
Z-0041-2020
Event Number
81419
Firm
Ohmeda Medical
FEI Number
1121732
Product Code
FMZ
Status
Terminated
Root Cause
Labeling design
Initiated
October 5, 2018
Terminated
June 2, 2020
Address
8880 Gorman Rd, Laurel, MD, 20723-5800

Description

Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Reason

GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.

Action

On October 5, 2018, the recalling firm notified affected customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to do the following: You can continue to use your Giraffe OmniBed systems in open Bed warmer mode and incubator mode. Care should be taken to achieve visible cleanliness during the cleaning process. Design of the Giraffe OmniBed allows for your facilities normal cleaning to be performed, following the step by step procedure to remove the canopy seal to reach visible debris accumulation. Contact your Biomedical department to perform the removal and reassembly of the original seal fasteners, per the attached addendum, prior to performing the cleaning procedure. Attention to cleaning technique by personnel involved in the cleaning process will reduce the likelihood of occurrence of the mentioned condition. "Carefully follow the additional instructions provided to ensure continued safe use of the device and place the addendum in your User Manual. "Always follow your facility guidelines for cleaning and infection control. GE Healthcare will provide a Giraffe OmniBed Cleaning & Care Guidelines poster for display by your system. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.