FDA Recall Terminated

Stationary Column TruSystem 7599, Material Number 1717020 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system.

Recall: Z-0041-2017 · Initiated July 10, 2016

Recall

Recall Number
Z-0041-2017
Event Number
75115
Firm
Trumpf Medical Systems, Inc.
FEI Number
3003184737
Product Code
JEA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 10, 2016
Posted
October 5, 2016
Terminated
March 2, 2017
Address
1046 Legrand Blvd, Charleston, SC, 29492-7672

Description

Stationary Column TruSystem 7599, Material Number 1717020 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system.

Reason

Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.

Action

Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543.

Distribution

Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland, Russia, Slovakia, and Switzerland

Quantity

84 units