FDA Recall
Terminated
Xoran MiniCAT CT Scanner ID ENT00001
Recall: Z-0041-2014
·
Initiated December 14, 2009
Recall
- Recall Number
- Z-0041-2014
- Event Number
- 66344
- Firm
- Xoran Technologies, Inc.
- FEI Number
- 3004198450
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- December 14, 2009
- Terminated
- September 11, 2014
- Address
- 5210 S State Rd, Ann Arbor, MI, 48108-7936
Description
Xoran MiniCAT CT Scanner ID ENT00001
Reason
Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
Action
The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.
Distribution
Worldwide Distribution - US nationwide
Quantity
320