FDA Recall Open, Classified

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Recall: Z-0040-2024 · Initiated August 30, 2023

Recall

Recall Number
Z-0040-2024
Event Number
92810
Firm
AB Sciex, LLC
FEI Number
3008935545
Product Code
DOP
Status
Open, Classified
Root Cause
Component design/selection
Initiated
August 30, 2023
Posted
October 5, 2023
Address
500 Old Connecticut Path, Framingham, MA, 01701-4574

Description

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Reason

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Action

An URGENT: RUO and MEDICAL DEVICE FIELD ACTION notification letter dated 7/20/23 was sent to customers. Actions to be taken by the Customer/User: SCIEX will provide a replacement nut for eligible sources and systems. Instructions will be provided to guide you through the replacement of the nut. The replacement parts with installation instructions are estimated to begin shipping in October 2023. SCIEX is currently notifying affected customers of the upcoming countermeasure. Upon availability, SCIEX will send you the replacement nut with instructions on how to install it on the ion source probe. We sincerely apologize for any inconvenience that this may cause you. SCIEX aims to provide you with products of the highest quality. If you have any questions regarding this notice, please contact https://sciex.com/aboutus/ contact-us or https://sciex.com/support/request-support.

Distribution

US Nationwide.

Quantity

14