FDA Recall Terminated

Xoran MiniCAT00202

Recall: Z-0040-2014 · Initiated July 27, 2010

Recall

Recall Number
Z-0040-2014
Event Number
66340
Firm
Xoran Technologies, Inc.
FEI Number
3004198450
Product Code
JAK
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
July 27, 2010
Terminated
September 11, 2014
Address
5210 S State Rd, Ann Arbor, MI, 48108-7936

Description

Xoran MiniCAT00202

Reason

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

Action

All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component. Xoran's Research & Development group is evaluating the current specification to see if the field specifications are appropriate and to ensure that proper testing protocols are communicated and are in place for use by 3 rd party testing organizations. In addition, Xoran has initiated CAPA # 18 0 that is looking at the source block with regard to evaluating the stability of the dose rating.

Distribution

Worldwide Distribution - Nationwide

Quantity

320