FDA Recall
Terminated
Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705
Recall: Z-0040-04
·
Initiated September 8, 2003
Recall
- Recall Number
- Z-0040-04
- Event Number
- 27251
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 8, 2003
- Posted
- October 15, 2003
- Terminated
- March 30, 2004
- Address
- 200 South Kraemer Blvd W, 337, Brea, CA, 92822
Description
Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705
Reason
Samples with high fat content will not count platelets accurately.
Action
Firm issued an advisory letter to users to retain as part of the Quality Manual for the instruments. It cautions to check specimens for interfering lipemia. Letter was sent 9/8/2003.
Distribution
United States and Canada
Quantity
451