Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.
Recall
- Recall Number
- Z-0039-2012
- Event Number
- 59316
- Firm
- Richard-Allan Scientific Company
- FEI Number
- 1831638
- Product Code
- IDO
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 12, 2011
- Posted
- October 12, 2011
- Terminated
- July 25, 2013
- Address
- 4481 Campus Dr, Kalamazoo, MI, 49008
Description
Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.
The firm has received two complaints of finger lacerations in operators of this device.
ThermoFisher Scientific sent "URGENT MEDICAL DEVICE CORRECTION, RE: Thermo Scientific HM355S-3 User Safety Advisory Notice" to all affected customers. The letter states the manufacturer identified a need to enhance the safety instructions and the description of the safety devices in the user manual. It mentions the HM3555S-3 User Manual has been modified to include more specific safe work practices and a new Quick -Reference Card summarizing the manual's safety enhancements and that the changes have been included on a CD that is enclosed with the letter. The firm requests confirmation of receipt of the new User Manual and Quick-Reference Card by October 31, 2011 to be sent by FAX to (269) 372-2674 For questions on this recall call the firms Technical Support representatives at 1-800-522-7270 08:00 AM- 4:30 PM.