FDA Recall
Terminated
Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.
Recall: Z-0038-2008
·
Initiated June 22, 2007
Recall
- Recall Number
- Z-0038-2008
- Event Number
- 38511
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- June 22, 2007
- Posted
- October 11, 2007
- Terminated
- April 25, 2012
- Address
- 810 Sharon Dr, Westlake, OH, 44145-1598
Description
Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.
Reason
Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.
Action
By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.
Distribution
Product was distributed throughout the United States and Canada.
Quantity
289 pieces