FDA Recall Terminated

Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.

Recall: Z-0038-2008 · Initiated June 22, 2007

Recall

Recall Number
Z-0038-2008
Event Number
38511
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Process design
Initiated
June 22, 2007
Posted
October 11, 2007
Terminated
April 25, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.

Reason

Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.

Action

By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.

Distribution

Product was distributed throughout the United States and Canada.

Quantity

289 pieces