FDA Recall Terminated

Datex-Ohmeda S/5 Aespire Anesthesia System 7100

Recall: Z-0037-06 · Initiated September 27, 2005

Recall

Recall Number
Z-0037-06
Event Number
33678
Firm
Datex-Ohmeda Inc
FEI Number
2183066
Product Code
BSZ
Status
Terminated
Root Cause
Other
Initiated
September 27, 2005
Posted
October 12, 2005
Terminated
April 1, 2007
Address
3030 Ohmeda Drive PO Box 7550, zip 53707-7550, Madison, WI, 53718

Description

Datex-Ohmeda S/5 Aespire Anesthesia System 7100

Reason

Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.

Action

Urgent Medical Device Correction letter dated 09/23/05 was sent to all customers. The letter describes the issues, affected device serial numbers and recommends that customers DO NOT USE the internal auxiliary oxygen flowmeter, if installed in their system, and to use alternate auxiliary oxygen sources, i.e. wall-mounted oxygen flow meter, in the interim. A GE service representative will contact the customers to schedule an upgrade, performed at no charge.

Distribution

CA, CO, FL, GA, IL, IN, MI, MS, NM, OK, PA, PR, SD, UT & WA. OUS to: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Canada, Caribbean, Czech Republic, Chile, China, Colombia, Dominican Republic, Ecuador, Finland, France, Germany, Ghana, Great Britain, Hong Kong, India, Indonesia, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Oman, Palestine, Peru, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Arab Emerites, Venezuela & Vietnam.

Quantity

367