Datex-Ohmeda S/5 Aespire Anesthesia System 7100
Recall
- Recall Number
- Z-0037-06
- Event Number
- 33678
- Firm
- Datex-Ohmeda Inc
- FEI Number
- 2183066
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 27, 2005
- Posted
- October 12, 2005
- Terminated
- April 1, 2007
- Address
- 3030 Ohmeda Drive PO Box 7550, zip 53707-7550, Madison, WI, 53718
Description
Datex-Ohmeda S/5 Aespire Anesthesia System 7100
Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
Urgent Medical Device Correction letter dated 09/23/05 was sent to all customers. The letter describes the issues, affected device serial numbers and recommends that customers DO NOT USE the internal auxiliary oxygen flowmeter, if installed in their system, and to use alternate auxiliary oxygen sources, i.e. wall-mounted oxygen flow meter, in the interim. A GE service representative will contact the customers to schedule an upgrade, performed at no charge.
CA, CO, FL, GA, IL, IN, MI, MS, NM, OK, PA, PR, SD, UT & WA. OUS to: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Canada, Caribbean, Czech Republic, Chile, China, Colombia, Dominican Republic, Ecuador, Finland, France, Germany, Ghana, Great Britain, Hong Kong, India, Indonesia, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Oman, Palestine, Peru, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Arab Emerites, Venezuela & Vietnam.
367