FDA Recall Terminated

TxRx 15CH Knee Coil 1.5T QED part number: Q7000050 Siemens Model # 10606524

Recall: Z-0034-2020 · Initiated August 20, 2019

Recall

Recall Number
Z-0034-2020
Event Number
83775
Firm
Quality Electrodynamics LLC
FEI Number
3007350713
Product Code
MOS
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 20, 2019
Terminated
May 26, 2020
Address
6655 Beta Dr, Ste 100, Mayfield Village, OH, 44143-2380

Description

TxRx 15CH Knee Coil 1.5T QED part number: Q7000050 Siemens Model # 10606524

Reason

Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.

Action

The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.

Distribution

Worldwide distribution - US in the states: NC OUS: Germany

Quantity

3 units