FDA Recall Terminated

Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007, 5105101008, 5109131010, 5111231011, 5114301012, 5102131101 & 5113091012. Manufactured by: Isopure Corp, Simpsonville, KY System is used to mix and distribute bicarbonate as a part of a clinical dialysis system.

Recall: Z-0033-2012 · Initiated May 19, 2011

Recall

Recall Number
Z-0033-2012
Event Number
59879
Firm
Isopure Corp
FEI Number
3013942511
Product Code
FJP
Status
Terminated
Root Cause
Software design
Initiated
May 19, 2011
Posted
October 11, 2011
Terminated
February 19, 2013
Address
141 Citizens Blvd, Simpsonville, KY, 40067-6548

Description

Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007, 5105101008, 5109131010, 5111231011, 5114301012, 5102131101 & 5113091012. Manufactured by: Isopure Corp, Simpsonville, KY System is used to mix and distribute bicarbonate as a part of a clinical dialysis system.

Reason

Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.

Action

Isopure sent a letter dated May 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Isopure devised an immediate solution which must be implemented at their faclity in conjunction with the disinfection procedure. Customers were asked to inform all appropriate personnel, clinical and technical, who may be involved in operating the system during the normal disinfection schedule. On 5/19/2011, the firm sent letters to their customers via US mail and email. They also telephoned their customers to notify them of the problem.

Distribution

Nationwide Distribution including AR, CT, GA, MA, ME, NJ, PA, TN and WV.

Quantity

11 units