Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.
Recall
- Recall Number
- Z-0032-2018
- Event Number
- 78093
- Firm
- St Jude Medical Inc.
- FEI Number
- 2017865
- Product Code
- NIK
- Status
- Terminated
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- August 28, 2017
- Terminated
- July 8, 2020
- Address
- 15900 Valley View Ct, Sylmar, CA, 91342-3577
Description
Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.
Worldwide Distribution
10,453 units