FDA Recall Terminated

Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.

Recall: Z-0032-2018 · Initiated August 28, 2017

Recall

Recall Number
Z-0032-2018
Event Number
78093
Firm
St Jude Medical Inc.
FEI Number
2017865
Product Code
NIK
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
August 28, 2017
Terminated
July 8, 2020
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577

Description

Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.

Reason

New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.

Action

Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.

Distribution

Worldwide Distribution

Quantity

10,453 units