FDA Recall Terminated

Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.

Recall: Z-0031-2022 · Initiated September 7, 2021

Recall

Recall Number
Z-0031-2022
Event Number
88673
Firm
NOX MEDICAL Keldnaholt Reykjavik Iceland
FEI Number
3007389703
Product Code
MNR
Status
Terminated
Root Cause
Finished device change control
Initiated
September 7, 2021
Terminated
May 1, 2023

Description

Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.

Reason

Error in the device firmware results in the inaccurate detection of device position.

Action

Written notification letter titled, "URGENT: Medical Device Recall" was sent to all consignees on 09/07/2021. The letter instructs the consignee to follow the following instructions: "Further distribution or use of the affected devices should cease immediately. Devices should be quarantined pending device firmware update, per the instructions below. Due to the potential of inaccurate position-related reporting of TST, TRT, AHI and REI; studies performed with the affected T3s devices should be reanalyzed to determine if this has clinical impact on the patient diagnoses. A Nox Medical technical representative will contact you within 5 business days to schedule a remote session (via videoconference or other means) to upgrade the device firmware. Nox Medical requests that you formally acknowledge receipt of this notice and return the attached Customer / Distributor - Acknowledgement and Receipt Form (Attachment 1). This recall notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If any affected devices which were in your possession (indicated in Attachment 2: Product Distribution and Information Table) have been transferred, please indicate this and provide as much detail as possible in the Acknowledgement and Receipt Form (Attachment 1)."

Distribution

US Nationwide distribution in the states of OH, IL, GA, IA, CA, PA, NM, MI, SD.

Quantity

128 units