Aligned Medical T and A Pack REF AMS11482 that contains the Cardinal Health Salem Sump PVC tubes. Product packaged in a convenient manner for use in a general clinical procedure.
Recall
- Recall Number
- Z-0030-2025
- Event Number
- 95306
- Firm
- Windstone Medical Packaging, Inc.
- FEI Number
- 1000125955
- Product Code
- LRO
- Status
- Open, Classified
- Root Cause
- Use error
- Initiated
- August 16, 2024
- Posted
- October 8, 2024
- Address
- 1602 4th Ave N, Billings, MT, 59101-1521
Description
Aligned Medical T and A Pack REF AMS11482 that contains the Cardinal Health Salem Sump PVC tubes. Product packaged in a convenient manner for use in a general clinical procedure.
Due to convenience kits containing a kit component that is under a recall action for improper use.
On 08/16/2024, the firm email an "URGENT: MEDICAL DEVICE PRODUCT CORRECTION" Letter to customers informing them that Cardinal Health is conducting a product correction for its Salem Sump PVC Tubes. These Sump PVC tubes were included in the Aligned Medical convenience kits. The correction for this Recall is an updated product label, an electronic instruction for use (eIFU). The updated labeling is to address improper use of the Sump PVC tubes which can lead to breakage of the Salem Sump Anti Reflux Valve (ARV) that increases risk to patients. Customers are instructed to: 1. Review your inventory for the affected kits, all lots are impacted 2. Communicate the change to the use instructions with all personnel that utilize Salem Sump" products contained within AMS kits. 3. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product correction. Your notification to your customers should include a copy of this notification letter. 4. Please complete and return the AMS Urgent Medical Device Product Correction Response Required form within 5 business days and return via email to [email protected] even if you have no affected product on hand. For questions or concerns, contact Aligned Medical Solutions at [email protected]
U.S. Nationwide distribution in the states of CA, MO, OH, VA, and WY.
40 packs