FDA Recall Terminated

L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.

Recall: Z-0030-2020 · Initiated October 10, 2018

Recall

Recall Number
Z-0030-2020
Event Number
83191
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FKX
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
October 10, 2018
Terminated
July 22, 2020
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.

Reason

Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.

Action

On 10/10/2018, and 10/15/2018, Urgent Product Recall notices were mailed to dialysis providers at medical centers. Customers were asked to do the following: 1. Locate and remove the affected product lot from your facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm for Service to arrange for return and credit. Center for Service can be reached at 888-229-0001. 3. If you purchased this product directly from the recalling firm, complete and return the enclosed customer reply form. 4. If you purchased this product from a distributor, please contact your distributor for return and credit. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a consumer-level recall of the affected product that you distributed to customers. Customers with additional questions are encouraged to contact: Corporate Product Surveillance at 800-437-5176, Clinical Helpline at 888-736-2543, and Renal Therapy Services at 800-553-6898. On 10/15/2018, Urgent Product Recall notices were mailed to Peritoneal Dialysis Patients who were asked to do the following: 1. Locate and remove the affected product lot from your supplies. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm to arrange for return and credit. HomeCare Services can be reached at 800-284-4060. 3. Complete and return the enclosed Home Patient Reply Form.

Distribution

U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.

Quantity

8640 eaches