FDA Recall Terminated

MRP-7000 and AIRIS Magnetic Resonance Imaging Systems, Software Versions: V7.0A to V7.0J.

Recall: Z-0029-2007 · Initiated August 16, 2006

Recall

Recall Number
Z-0029-2007
Event Number
36224
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
August 16, 2006
Posted
October 12, 2006
Terminated
April 13, 2012
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

MRP-7000 and AIRIS Magnetic Resonance Imaging Systems, Software Versions: V7.0A to V7.0J.

Reason

Image orientation error. When a 3D Maximum Intensity Projection (MIP) image data set is transferred from the MRI system to a computer workstation via the DICOM protocol, the anatomical markers do not change between the images as they rotate.

Action

On 8/16/2006, Hitachi issued two (2) Device Correction letters. The first letter was addressed to the five (5) customers whose system units are currently ''de-installed''. This letter informs the customers of the problem and asks that the customer(s) please contact the firm''s Customer Service Center on their toll-free tel. # 1-800-800-4925, in order to schedule a device correction should the customer ever re-activate/install their system(s). The second letter, issued to the customers with activated MRP-7000 and/or AIRIS MRI systems, informs the customers of the problem, and states that a HMSA Service (person) will be contacting the customer in order to schedule a device correction which should address the problem.

Distribution

Nationwide and Puerto Rico.

Quantity

348 units