FDA Recall Terminated

BBL (tm) MGIT (tm) Mycobacteria Growth Indicator Tube, part # 245122, packaged in 100 ct cartons and labeled in part Product Usage: The Mycobacteria Growth Indicator Tube supplemented with Bactec MGIT Growth Supplement and BBL MGIT PANTA antibiotic mixture is intended for the detection and recovery of mycobacteria using the BACTEC MGIT 960 and BACTEC MGIT 320 Systems. Acceptable specimen types are digested and decontaminated clinical specimens (except urine), and sterile body fluids (except blood).

Recall: Z-0028-2014 · Initiated September 9, 2013

Recall

Recall Number
Z-0028-2014
Event Number
66325
Firm
Becton Dickinson & Co. BD Diagnostic Systems
FEI Number
1119779
Product Code
MDB
Status
Terminated
Root Cause
Error in labeling
Initiated
September 9, 2013
Posted
October 19, 2013
Terminated
October 29, 2013
Address
7 Loveton Circle, Sparks, MD, 21152-0999

Description

BBL (tm) MGIT (tm) Mycobacteria Growth Indicator Tube, part # 245122, packaged in 100 ct cartons and labeled in part Product Usage: The Mycobacteria Growth Indicator Tube supplemented with Bactec MGIT Growth Supplement and BBL MGIT PANTA antibiotic mixture is intended for the detection and recovery of mycobacteria using the BACTEC MGIT 960 and BACTEC MGIT 320 Systems. Acceptable specimen types are digested and decontaminated clinical specimens (except urine), and sterile body fluids (except blood).

Reason

An in-vitro diagnostic reagent may be labeled with an incorrect barcode, which when scanned on associated diagnostic instrument may cause a false negative result.

Action

BD sent an Urgent Product Recall letter dated September 2013 to customers identified as microbiology lab managers. The letter identified the affected product, problem and actions to be taken. The letter advised that referenced lot numbers may contain an overlabel barcode that may begin to lift, exposing the original incorrect barcode pictured in the notification. Customers were instructed to complete the attached form and fax to Regulatory Compliance 410-316-4258. For further assistance regarding replacements contact BD Customer Service Department at 1-800-675-0908 and all other inquiries contact BD Technical Services Department at 1-800-638-8663.

Distribution

USA Nationwide Disatribution in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, IL, KY, MA, ME, MN, MO, NC, ND, NJ, NV, NY, OH, PA, TX, UT, VA, WA, WI

Quantity

187,000 tubes