Lactate Disposable Membrane Caps Product Number: 0018108400
Recall
- Recall Number
- Z-0027-04
- Event Number
- 27163
- Firm
- Instrumentation Laboratory Co.
- FEI Number
- 1217183
- Product Code
- LNW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 28, 2003
- Posted
- October 15, 2003
- Terminated
- May 23, 2012
- Address
- 113 Hartwell Avenue, Lexington, MA, 02173
Description
Lactate Disposable Membrane Caps Product Number: 0018108400
Lactate Disposable Membrane Caps linearity was determined to be significantly biased and could potentially cause erroneous patient results above 6.0 mmol/L.
Instrumentation Laboratories notified two (2) U.S. direct customers on August 28, 2003 by telephone, regarding Lot N30601. Follow-up phoning emphasizing significance of linearity problem (and additional involved lot N307 - for 1 affected customer) performed on September 3, 2003. Phone call followed with faxback acknowledging notification and disposition of remaining product lot(s). IL service also dispatched to customer sites on September 4, 2003. International Distributor Notices sent to IL Europe and IL Mexico on September 4, 2003. Product Information Bulletins circulated on September 3, 2003, to IL personnel worldwide.
IL, NY International: Italy, Mexico
287 boxes