FDA Recall Terminated

Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.

Recall: Z-0023-05 · Initiated August 17, 2004

Recall

Recall Number
Z-0023-05
Event Number
29864
Firm
Stryker Medical
FEI Number
1831750
Product Code
FPO
Status
Terminated
Root Cause
Other
Initiated
August 17, 2004
Posted
October 20, 2004
Terminated
October 19, 2004
Address
6300 Sprinkle Rd, Kalamazoo, MI, 49001

Description

Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.

Reason

Locking pins may be installed backward, and cot may drop from highest to lowest position when the users are not prepared for the drop.

Action

Service representatives visited each customer, inspected the cots, and reinstalled the lock pins, if required, beginning on August 10, 2004.

Distribution

Alabama, California, Florida, Kansas, Montana, Oklahoma, Texas and Virginia.

Quantity

15