FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Recall
- Recall Number
- Z-0022-2025
- Event Number
- 95326
- Firm
- Qualigen Inc
- FEI Number
- 3003355251
- Product Code
- JLW
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 26, 2024
- Posted
- October 8, 2024
- Address
- 2042 Corte Del Nogal, Ste B, Carlsbad, CA, 92011-1438
Description
FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
On 8/26/24, correction notices were emailed to customers who were asked to do the following: 1) Dispose of affected TSH calibration cards. 2) Use TSH Calibration Card Lot Number 2405017-1R sent with correction notices. 3) Run TSH Quality Control samples. 4) Review and retest any affected TSH results that may have been higher than expected. 5) Notify all relevant parties within your organization or any organization that received the affected devices. 6) Complete and return response forms via email to [email protected] If you have any questions regarding this matter, contact System Support Specialists at [email protected] or (877) 709-2169, option 2.
US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland