FDA Recall Terminated

GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company

Recall: Z-0021-2007 · Initiated August 10, 2006

Recall

Recall Number
Z-0021-2007
Event Number
36057
Firm
Datex Ohmeda, Inc
FEI Number
2183066
Product Code
BSZ
Status
Terminated
Root Cause
Other
Initiated
August 10, 2006
Posted
November 7, 2006
Terminated
February 24, 2008
Address
Po Box 7550, Madison, WI, 53707

Description

GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company

Reason

Vaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.

Action

Firm mailed recall letters on August 10, 2006, to consignees telling them what to do if a failure occurred during system checkout or during use, as a short term solution. Later, GE representatives will make visits to each site for modification of the interface valve on each device, as the long term solution.

Distribution

AK, AL, CA, FL, KS, KY, MD, MN, MO, NC, NE, SC, SD, TX, VA, WI,

Quantity

471