GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company
Recall
- Recall Number
- Z-0021-2007
- Event Number
- 36057
- Firm
- Datex Ohmeda, Inc
- FEI Number
- 2183066
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 10, 2006
- Posted
- November 7, 2006
- Terminated
- February 24, 2008
- Address
- Po Box 7550, Madison, WI, 53707
Description
GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company
Vaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.
Firm mailed recall letters on August 10, 2006, to consignees telling them what to do if a failure occurred during system checkout or during use, as a short term solution. Later, GE representatives will make visits to each site for modification of the interface valve on each device, as the long term solution.
AK, AL, CA, FL, KS, KY, MD, MN, MO, NC, NE, SC, SD, TX, VA, WI,
471