FDA Recall Open, Classified

OPTETRAK Comprehensive Knee System including OPTETRAK All-polyethylene UNI Tibial Components, Product Lines, 252-12-XX, 252-13-XX, 252-22-XX, 252-23-XX. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

Recall: Z-0020-2022 · Initiated August 30, 2021

Recall

Recall Number
Z-0020-2022
Event Number
88570
Firm
Exactech, Inc.
FEI Number
1038671
Product Code
HRY
Status
Open, Classified
Root Cause
Process control
Initiated
August 30, 2021
Posted
October 4, 2021
Address
2320 Nw 66th Ct, Gainesville, FL, 32653-1630

Description

OPTETRAK Comprehensive Knee System including OPTETRAK All-polyethylene UNI Tibial Components, Product Lines, 252-12-XX, 252-13-XX, 252-22-XX, 252-23-XX. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

Reason

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Action

Exactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year shelf life not packaged in EVOH/Nylon bags. A communication to healthcare professionals should follow. Instructions to recipients included to immediately identify and quarantine any affected product in inventory, notify customers down the line if further distributed, and complete and return the Recall Inventory Response Form to Exactech via email at [email protected]. Exactech expanded their recall and sent an updated notification to US agents and distributors on 02/07/2022. On about 02/08/2022, a letter to surgeons and hospitals was sent. The notifications indicated that all non-conforming devices were included in the recall, to cease use of those devices, and work with sales agents to return any inventory. The letters to surgeons and hospitals also advised physicians regarding follow up with implanted patients.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TUNISIA, TURKEY, VENEZUELA .

Quantity

2,774 devices