FDA Recall
Terminated
Proteus 235 Proton Therapy System for cancer treatment.
Recall: Z-0020-2015
·
Initiated March 27, 2014
Recall
- Recall Number
- Z-0020-2015
- Event Number
- 68846
- FEI Number
- 3000256071
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 27, 2014
- Posted
- October 29, 2014
- Terminated
- December 10, 2014
- Address
- Ion Beam Applications S.A. Chemin du Cyclotron, 3 Louvain La Neuve Belgium
Description
Proteus 235 Proton Therapy System for cancer treatment.
Reason
While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.
Action
A User Notice was sent to direct accounts in/about March/April 2014 to notify them of the issue and provide instructions to avoid the identified potential risk to patients.
Distribution
Worldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.
Quantity
8