FDA Recall
Terminated
Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.
Recall: Z-0019-2019
·
Initiated September 4, 2018
Recall
- Recall Number
- Z-0019-2019
- Event Number
- 80986
- Firm
- Leica Microsystems, Inc.
- FEI Number
- 3000210133
- Product Code
- KIM
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- September 4, 2018
- Terminated
- July 10, 2020
- Address
- 1700 Leider Ln, Buffalo Grove, IL, 60089-6622
Description
Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.
Reason
Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.
Action
The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction.
Distribution
Distribution in US state of MO.
Quantity
1 unit