FDA Recall Terminated

Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.

Recall: Z-0019-2019 · Initiated September 4, 2018

Recall

Recall Number
Z-0019-2019
Event Number
80986
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
KIM
Status
Terminated
Root Cause
Process design
Initiated
September 4, 2018
Terminated
July 10, 2020
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.

Reason

Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.

Action

The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction.

Distribution

Distribution in US state of MO.

Quantity

1 unit