FDA Recall Open, Classified

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Recall: Z-0016-2026 · Initiated August 29, 2025

Recall

Recall Number
Z-0016-2026
Event Number
97535
Firm
Getinge Disinfection Ab Ljungadalsgatan
FEI Number
3000231474
Product Code
MEC
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 29, 2025
Posted
October 1, 2025
Address
11 Vaxjo Sweden

Description

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Reason

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Action

Getinge notified consignees on about 08/29/2025 via letter. Consignees were instructed that Getinge service technicians will be proactively reaching out to ensure affected devices are calibrated appropriately. In the meantime, Getinge has developed a verification procedure to ensure that the circulation pressure in standby mode is correctly calibrated to 0.0 kPa. If the value deviates, recalibration by a qualified technician is required. The verification of the circulation pressure shall be performed by the operator or Getinge technician. If the value deviates from 0.0 kPa out of allowed tolerances (1 kPa), stop using the device immediately and work with your Getinge service technician to arrange recalibration. Consignees were also instructed to ensure all caregivers and users of the affected product are made aware of the recall and to complete and return the provided response form. ***UPDATE 10/16/2025: Firm notified FDA of an expansion to include additional OUS units and one additional complaint. ***UPDATE 03/02/2026: Firm notified FDA of an expansion to include additional OUS units and one additional complaint.

Distribution

Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.

Quantity

902 units