FDA Recall Terminated

Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical, Long Term Care Division, 5001 Joerns Drive, Stevens Point, WI 54481 USA.

Recall: Z-0016-2007 · Initiated January 1, 2006

Recall

Recall Number
Z-0016-2007
Event Number
35816
Firm
Sunrise Medical CCG, Inc.
FEI Number
2182305
Product Code
FNL
Status
Terminated
Root Cause
Other
Initiated
January 1, 2006
Posted
October 5, 2006
Terminated
April 9, 2007
Address
5501 Joerns Dr, Stevens Point, WI, 54481-5011

Description

Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical, Long Term Care Division, 5001 Joerns Drive, Stevens Point, WI 54481 USA.

Reason

The control box used in the actuator systems appears to have a potential to become stuck in one position resulting a non-functioning bed.

Action

The objective is to replace all of these systems. Phone contact of affected facilities began in January 2006. All facilities to be contacted by June 23, 2006. All products will be corrected at the location of the end-user by June 30, 2006.

Distribution

Nationwide and Canada.

Quantity

6,921