FDA Recall Terminated

Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Number 25660996 (2) DeRoyal (R) Adult Anesthesia Circuit, REF 86-001781, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 (3) DeRoyal (R) Latex Safe Anesthesia Kit, REF 86-001641, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25573171, 25765667, 25834853, 25834861 (4) DeRoyal (R) Custom Anesthesia Kit, REF 86-001619, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Number 25573155 (5) DeRoyal (R) Anesthesia Kit, REF 86-001669, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25313484, 25702324, 25977977 Intended Use: Varies by kit; an anesthesia breathing circuit is a device that is intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Y-piece.

Recall: Z-0015-2012 · Initiated August 23, 2011

Recall

Recall Number
Z-0015-2012
Event Number
59854
Firm
Deroyal Industries, Inc. Lafollette
FEI Number
3005011024
Product Code
OGD
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 23, 2011
Posted
October 6, 2011
Terminated
January 30, 2013
Address
1501 East Central Ave La, Follette, TN, 37766-2892

Description

Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Number 25660996 (2) DeRoyal (R) Adult Anesthesia Circuit, REF 86-001781, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 (3) DeRoyal (R) Latex Safe Anesthesia Kit, REF 86-001641, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25573171, 25765667, 25834853, 25834861 (4) DeRoyal (R) Custom Anesthesia Kit, REF 86-001619, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Number 25573155 (5) DeRoyal (R) Anesthesia Kit, REF 86-001669, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25313484, 25702324, 25977977 Intended Use: Varies by kit; an anesthesia breathing circuit is a device that is intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Y-piece.

Reason

The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the patient.

Action

The firm, DeRoyal, sent a "URGENT: MEDICAL DEVICE RECALL" letter dated September 13, 2011 to its customers. The letter describes product, problem, and actions to be taken. Customers are instructed to identify inventory being recalled and place in quarantine to prevent further use. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Customers were asked to complete the attached Recall Response Spreadsheet and return as per the instructions. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Distributors were instructed to perform a sub-recall to their customers. Customers with questions should call 865-362-1037 between 9am and 5pm EST or email [email protected].

Distribution

Nationawide Distribution USA - including the states of: CA, FL, GA, NJ, and WA.

Quantity

940 units