Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic.
Recall
- Recall Number
- Z-0015-2008
- Event Number
- 39389
- Firm
- Thermo Fisher Scientific
- FEI Number
- 2550000
- Product Code
- JIF
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 24, 2007
- Posted
- October 4, 2007
- Terminated
- December 18, 2007
- Address
- 171 Industry Dr, Pittsburgh, PA, 15275-1015
Description
Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic.
Method is susceptible to positive interference from endogenous ALT in patient samples.
The firm has updated the Limitations section of the Ammonia Reagent Package Insert and circulated a Technical Bulletin dated July 24, 2007 to all customers that have purchased Ammonia Reagent that is still within expiration date. A Notification of Change has been introduced and shall be placed within subsequent Ammonia Reagent Kits with a copy of the updated package insert; the notification of change informs customers that a change has been made to the package insert. Notifications of Change are placed within kits for 6 months from the date of issue.
Worldwide, including USA, Ecuador, and Canada.
6,882 units