FDA Recall
Terminated
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Recall: Z-0014-2018
·
Initiated July 28, 2016
Recall
- Recall Number
- Z-0014-2018
- Event Number
- 78130
- Firm
- Handicare AB Maskinvagen
- FEI Number
- 3009481053
- Product Code
- FSA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 28, 2016
- Terminated
- December 17, 2018
- Address
- 17 Lulea Sweden
Description
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Reason
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Action
The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.
Distribution
OH, PA, MD
Quantity
9 units