FDA Recall Terminated

PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Recall: Z-0013-2018 · Initiated September 8, 2017

Recall

Recall Number
Z-0013-2018
Event Number
78096
Firm
Handicare Usa Inc
FEI Number
3007802293
Product Code
FSA
Status
Terminated
Root Cause
Component design/selection
Initiated
September 8, 2017
Posted
September 15, 2017
Terminated
December 17, 2018
Address
10888 Metro Ct, Maryland Heights, MO, 63043-2413

Description

PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Reason

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

Action

The firm initiated their recall by email on 09/08/2017. The consignee was warned that the lift straps may wear prematurely and they would be contacted by a representative and the units would be replaced.

Distribution

Us and Canada

Quantity

8 units