Nuclear medicine systems: Gemini 16, Model 4535 679 28961. Gemini GXL, Models # 4535 679 71891, #4535 679 75681, #4535 679 75691. Precedence, Model #2169 -3000A, #2169-3001A, #2169-3002A, and 2169-3003A. PET/CT Computed tomography systems.
Recall
- Recall Number
- Z-0013-2007
- Event Number
- 37567
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 6, 2006
- Posted
- September 26, 2007
- Terminated
- October 5, 2010
- Address
- 595 Miner Road, Cleveland, OH, 44143-2131
Description
Nuclear medicine systems: Gemini 16, Model 4535 679 28961. Gemini GXL, Models # 4535 679 71891, #4535 679 75681, #4535 679 75691. Precedence, Model #2169 -3000A, #2169-3001A, #2169-3002A, and 2169-3003A. PET/CT Computed tomography systems.
A potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator.
The customers were sent Public Safety Notification November 3, 2006, indicating there was a problem with the software. However, since their unit was recently upgraded the problem did not impact their unit.
Worldwide, including USA, Australia, Canada, China, England, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Netherlands, Poland, Puerto Rico, Spain, and Taiwan.
70