FDA Recall
Terminated
EBI Trochanteric 11mm Long Nail.
Recall: Z-0012-06
·
Initiated June 24, 2005
Recall
- Recall Number
- Z-0012-06
- Event Number
- 32813
- FEI Number
- 2242816
- Product Code
- JDS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 24, 2005
- Posted
- October 7, 2005
- Terminated
- July 3, 2006
- Address
- EBI, L.P. 100 Interpace Pkwy, Parsippany, NJ, 07054-1149
Description
EBI Trochanteric 11mm Long Nail.
Reason
Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.
Action
The firm sent out via e-mail notification to their field representatives on 6/27/2005 to identify, remove and return to EBI all affected product from their inventories. One direct customer, Dr. Maylack was sent a recall letter on 7/7/2005.
Distribution
There are 9 sales representatives that are instructed to remove and return the EBI nail from nationwide customers.
Quantity
102 nails