FDA Recall Terminated

Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.

Recall: Z-0012-05 · Initiated September 3, 2004

Recall

Recall Number
Z-0012-05
Event Number
29988
Firm
General Electric Med Systems LLC
FEI Number
2126677
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
September 3, 2004
Posted
October 7, 2004
Terminated
July 24, 2006
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.

Reason

GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.

Action

A letter dated August 5, 2004 was released to customers describing the issue and recommending that users pay special attention to any volume curve which appears to be abnormal or any EF values that appear to be lower than expected, and not to rely soley on this value for diagnosis purposes until they recieve the SW patch. A SW patch will be distributed shortly to all affected systems.

Distribution

Nationwide throughout the United States and worldwide.

Quantity

853 units