Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.

Recall: Z-0012-04 · Initiated May 8, 2003

Recall

Recall Number
Z-0012-04
Event Number
26231
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Status
Terminated
Root Cause
Other
Initiated
May 8, 2003
Posted
October 15, 2003
Terminated
December 9, 2003
Address
9115 Hague Road, Indianapolis, IN, 46250-0457

Description

Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.

Reason

Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.

Action

Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.

Distribution

United States

Quantity

478