FDA Recall Open, Classified

ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D

Recall: Z-0011-2022 · Initiated September 3, 2021

Recall

Recall Number
Z-0011-2022
Event Number
88501
Firm
Welch Allyn Inc Mortara
FEI Number
2183461
Product Code
DPS
Status
Open, Classified
Root Cause
Software design
Initiated
September 3, 2021
Address
7865 N 86th St, Milwaukee, WI, 53224-3431

Description

ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D

Reason

The devices malfunction under specific operator workflows.

Action

The firm notified its consignees on 9/3/2021 by letter via U PS or via USPS first class for those with P.O. Box mailing addresses. The letter notified customers of the issue, provided clinical mitigations and advised them not to utilize specific workflows until their software is updated.

Distribution

Distribution was nationwide, including Puerto Rico. Foreign distribution was made to Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Czechia, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guatemala, Hongkong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Trinidad and Tobago, Turkey, United Arab Emirates, and United Kingdom.

Quantity

14,630 units