FDA Recall Terminated

MAQUET HLM Tubing set with Bioline Coating (BEQ-TOP 21400 with 1/2 x 1/2 HG 0286 Connector; Device Part Number BEQ-TOP 21400

Recall: Z-0011-2011 · Initiated September 7, 2010

Recall

Recall Number
Z-0011-2011
Event Number
56728
Firm
Maquet Inc.
FEI Number
3008355164
Product Code
DWE
Status
Terminated
Root Cause
Device Design
Initiated
September 7, 2010
Posted
October 6, 2010
Terminated
September 25, 2012
Address
45 Barbour Pond Drive, Wayne, NJ, 07470

Description

MAQUET HLM Tubing set with Bioline Coating (BEQ-TOP 21400 with 1/2 x 1/2 HG 0286 Connector; Device Part Number BEQ-TOP 21400

Reason

Tubing packs may contain defective 1/2 inch x 1/2 inch tubing connectors. Leaks at the shank of the connector may result in air aspiration in the venous line.

Action

Maquet Getigne Group issued Product Recall/Urgent Medical Device Corrective Action letters dated September 7, 2010 to customers, identifying the affected product, the issue prompting the recall, and actions to be taken by customers.Customers are to check their inventory for affected product. Customers with affected product are to contact Laura Olilver at 201-995-8849 for an RMA number and shipment instructions. All firms are to coomplete and return the Corrective Action Response Form. Maquet Customer Service can bre contacted at 800-777-4222.

Distribution

Worldwide Distribution: USA State of IL and country of the Netherlands.

Quantity

22 items