FDA Recall
Terminated
Medtronic MiniMed Paradigm Model 712
Recall: Z-0010-05
·
Initiated September 14, 2004
Recall
- Recall Number
- Z-0010-05
- Event Number
- 30059
- Firm
- Medtronic MiniMed
- FEI Number
- 3003166194
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 14, 2004
- Terminated
- March 3, 2005
- Address
- 18000 Devonshire St, Northridge, CA, 91325-1219
Description
Medtronic MiniMed Paradigm Model 712
Reason
Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.
Distribution
Nationwide & Spain
Quantity
10,600