iFusion, Device Listing Number D203271. Product Usage: The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. The iCam takes digital images of the posterior and external structures of the eye without the use of a mydriatic agent and is intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. iCam provides images only and does not provide any diagnostic, pathological analysis or classification of ocular health or disease. AND iVue with Normative Database (K121739) - The iVue is a noncontact, high resolution optical coherence tomography system intended for in vivo imaging, axial cross-sectional, three-dimensional imaging and measurement of anterior and posterior. ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects.
Recall
- Recall Number
- Z-0009-2018
- Event Number
- 77776
- Firm
- Optovue, Inc.
- FEI Number
- 3005950902
- Product Code
- HKI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 27, 2017
- Terminated
- April 12, 2021
- Address
- 2800 Bayview Dr, Fremont, CA, 94538-6518
Description
iFusion, Device Listing Number D203271. Product Usage: The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. The iCam takes digital images of the posterior and external structures of the eye without the use of a mydriatic agent and is intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. iCam provides images only and does not provide any diagnostic, pathological analysis or classification of ocular health or disease. AND iVue with Normative Database (K121739) - The iVue is a noncontact, high resolution optical coherence tomography system intended for in vivo imaging, axial cross-sectional, three-dimensional imaging and measurement of anterior and posterior. ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects.
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Optovue sent an Urgent Medical Device Software Field Correction letter dated July 27, 2017, via FedEx mail. Customers were Instructed to discontinue use of the Vault Mapping software, to complete and return the recall response form, and to provide remote access to units to an Optovue representative so that new software can be downloaded remotely. Questions can be addressed by calling Optovue at 1-866-941-9240.
US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, & WV.
326 devices