FDA Recall Terminated

Synchron LX Systems Version 4.5 Operating Software

Recall: Z-0009-06 · Initiated September 6, 2005

Recall

Recall Number
Z-0009-06
Event Number
33453
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
September 6, 2005
Posted
October 6, 2005
Terminated
January 26, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Synchron LX Systems Version 4.5 Operating Software

Reason

Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.

Action

Customers were notified by letter dated September 6, 2005 with work around instruction provided in the letter.

Distribution

Nationwide and Canada

Quantity

868