FDA Recall Open, Classified

Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

Recall: Z-0004-2025 · Initiated August 30, 2024

Recall

Recall Number
Z-0004-2025
Event Number
95303
Firm
Mercury Enterprises, Inc. dba Mercury Medical
FEI Number
1024404
Product Code
BTL
Status
Open, Classified
Root Cause
Process control
Initiated
August 30, 2024
Posted
October 2, 2024
Address
11300 49th St N, Clearwater, FL, 33762-4807

Description

Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

Reason

Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.

Action

Mercury Medical notified customers on 09/04/2024 via letter. Consignees who are medical facilities were instructed to immediately review inventory and quarantine any affected units on hand, contact customer service to arrange return of units, and complete the return response form. Consignees who are distributors were instructed to notify customers if further distributed, obtain a response form from those customers, arrange for any return of affected units, and complete the return response form.

Distribution

US distribution to MI, FL, SC, PA, IN. International distribution to Albania.

Quantity

1,300 units