Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.
Recall
- Recall Number
- Z-0003-2009
- Event Number
- 53158
- Firm
- GE Medical Systems, LLC
- FEI Number
- 2126677
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 14, 2008
- Posted
- September 22, 2009
- Terminated
- December 18, 2011
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.
GE Healthcare has recently become aware of an error in the generator software on the Revolution XR/d configured with HP 8200 PC consoles that reports a mAs reading that is higher than is actually exposed to patients.
Consignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology, and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, and Contact Information.
Worldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, ID, IL, IN, LA, MA, MI, MO, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA, and WI and countries of Canada, China, Japan, Korea, and Taiwan.
83 (63 USA, 20 OUS)