FDA Recall Terminated

Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.

Recall: Z-0003-2009 · Initiated February 14, 2008

Recall

Recall Number
Z-0003-2009
Event Number
53158
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
KPR
Status
Terminated
Root Cause
Software design
Initiated
February 14, 2008
Posted
September 22, 2009
Terminated
December 18, 2011
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.

Reason

GE Healthcare has recently become aware of an error in the generator software on the Revolution XR/d configured with HP 8200 PC consoles that reports a mAs reading that is higher than is actually exposed to patients.

Action

Consignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology, and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, and Contact Information.

Distribution

Worldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, ID, IL, IN, LA, MA, MI, MO, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA, and WI and countries of Canada, China, Japan, Korea, and Taiwan.

Quantity

83 (63 USA, 20 OUS)