FDA Recall Terminated

FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.

Recall: Z-0002-2008 · Initiated February 22, 2007

Recall

Recall Number
Z-0002-2008
Event Number
38369
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
LXG
Status
Terminated
Root Cause
Labeling design
Initiated
February 22, 2007
Posted
October 16, 2007
Terminated
April 26, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.

Reason

Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.

Action

A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.

Distribution

Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.

Quantity

9 in the U.S.; 1 in Canada