CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Recall
- Recall Number
- Z-0001-2017
- Event Number
- 75118
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 3000204839
- Product Code
- DRY
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 11, 2015
- Terminated
- October 26, 2016
- Address
- 125 Blue Ball Rd, Elkton, MD, 21921-5315
Description
CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
Terumo (Planned Action) plans to execute the following steps as part of the correction activity: Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter. For further questions, please call (410) 392-7277.
Worldwide Distribution - USA (nationwide) Distribution.
140,000 units