FDA Recall Terminated

CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Recall: Z-0001-2017 · Initiated December 11, 2015

Recall

Recall Number
Z-0001-2017
Event Number
75118
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
3000204839
Product Code
DRY
Status
Terminated
Root Cause
Other
Initiated
December 11, 2015
Terminated
October 26, 2016
Address
125 Blue Ball Rd, Elkton, MD, 21921-5315

Description

CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Reason

Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Action

Terumo (Planned Action) plans to execute the following steps as part of the correction activity: Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter. For further questions, please call (410) 392-7277.

Distribution

Worldwide Distribution - USA (nationwide) Distribution.

Quantity

140,000 units