FDA PMA 30-Day Notice Accepted 🇺🇸 United States

PMA: P970061 · Supplement: S010 · Decision Mar 27, 2001
Classifications
0
FEI Numbers
0
Registration Numbers
0

Basic Information

Trade Name
RADIUS(TM) CORONARY STENT WITH DELIVERY SYSTEM
PMA Number
P970061
Supplement Number
S010
Advisory Committee
Cardiovascular
Decision
30-Day Notice Accepted
Decision Code
OK30
Decision Date
March 27, 2001
Date Received
February 27, 2001
Supplement Type
30-Day Notice
Supplement Reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review
N

Advisory Committee Statement

CHANGE FROM AN INTERNAL PARTICULATE MAXIMUM LEVEL TO AN INTERNAL PARTICULATE ACTION LEVEL OF 75,000 PARTICLES >=10 MICRONS FOR THE FINISHED PRODUCT.