FDA PMA
FDA Class 3
30-Day Notice Accepted
🇺🇸 United States
Catheter, Coronary, Atherectomy
PMA: P900056
·
Supplement: S168
·
Decision Dec 1, 2017
Classifications
1
FEI Numbers
22
Registration Numbers
22
Basic Information
- Device Name
- Catheter, Coronary, Atherectomy
- Trade Name
- Rotablator Rotational Atherectomy System
- PMA Number
- P900056
- Supplement Number
- S168
- Device Class
- FDA Class 3
- Product Code
- MCX
- Generic Name
- CATHETER, CORONARY, ATHERECTOMY
- Medical Specialty
- Unknown
- Advisory Committee
- Cardiovascular
- Decision
- 30-Day Notice Accepted
- Decision Code
- OK30
- Decision Date
- December 1, 2017
- Date Received
- October 30, 2017
- Supplement Type
- 30-Day Notice
- Supplement Reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Expedited Review
- N
Advisory Committee Statement
Update the BSC2000-2 cycle conducted in Chamber 5 at the BSC Coventry, RI facility with new Programmable Logic Controller (PLC) software, a replacement vacuum pump with an additional vacuum pump booster, and a new validation documentation structure used to support the system.
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| MCX | Catheter, Coronary, Atherectomy | FDA class 3 | Unknown |