FDA PMA
FDA Class 3
30-Day Notice Accepted
🇺🇸 United States
Catheter, Coronary, Atherectomy
PMA: P900056
·
Supplement: S035
·
Decision Feb 11, 2000
Classifications
1
FEI Numbers
22
Registration Numbers
22
Basic Information
- Device Name
- Catheter, Coronary, Atherectomy
- Trade Name
- ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM
- PMA Number
- P900056
- Supplement Number
- S035
- Device Class
- FDA Class 3
- Product Code
- MCX
- Generic Name
- CATHETER, CORONARY, ATHERECTOMY
- Medical Specialty
- Unknown
- Advisory Committee
- Cardiovascular
- Decision
- 30-Day Notice Accepted
- Decision Code
- OK30
- Decision Date
- February 11, 2000
- Date Received
- January 21, 2000
- Supplement Type
- 30-Day Notice
- Supplement Reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Expedited Review
- N
Advisory Committee Statement
The 30-day notice requested two changes;the addition of an inspection of the advancer short drive shaft following the wet test and an additional glue application during assembly of the large hypo tube and liner components.
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| MCX | Catheter, Coronary, Atherectomy | FDA class 3 | Unknown |